Bilateral Impedance Rheography in Deep Venous ThrombosisRole of Respiratory Maneuvers and Saphenous Occlusion
Alan B. Gazzaniga, MD;
Robert H. Bartlett, MD;
Judy B. Shobe, RN
AMA Arch Surg. 1973;106(6):835-837.
Abstract
Bilateral impedance rheography with a respiratory maneuver has proven useful in the diagnosis of limb deep venous thrombosis. However, both false positive and false negative results have occurred with this method.
Using an inspiratory maneuver, 22 of 50 "normal" volunteer limbs had false positive results. When the Valsalva maneuver was used, all limb impedance responses reverted to normal. In 64 patients 23 had false positive tests with the inspiratory maneuver but were normal when the Valsalva maneuver was used. Eight patients with phlebographic proof of deep venous thrombosis had positive results with either respiratory maneuver. False negative results in five patients were reverted by placing tourniquets snugly around the thigh. Impedance rheography remains a useful screening procedure for the detection of limb deep venous thrombosis.
Author Affiliations
Orange, Calif
From the Department of Surgery, University of California, Irvine, Orange County Medical Center, Orange, Calif.
Footnotes
Accepted for publication Jan 22, 1973.
Reprint requests to Department of Surgery, University of California, Irvine, Orange County Medical Center, Orange, CA 92668 (Dr. Gazzaniga).
CiteULike Connotea Del.icio.us Digg Reddit Technorati Twitter
What's this?
THIS ARTICLE HAS BEEN CITED BY OTHER ARTICLES
Thrombophlebitis
Kerstein
ANGIOLOGY 1977;28:228-234.
Bedside Screening for Venous Thrombosis Using Occlusive Impedance Phlebography
Brownell Wheeler et al.
ANGIOLOGY 1975;26:199-210.
125I-Fibrinogen Uptake and Bilateral Impedance Rheography: Diagnosis of Postoperative Venous Thrombosis
Gazzaniga et al.
Arch Surg 1974;108:66-68.
ABSTRACT
|