Chronic testing of a pacemaker that needs recharging only once every four years
G. F. Tyers, R. R. Brownlee, H. C. Hughes Jr, N. J. Manley, I. N. Gorman and J. A. Waldhausen
Since 1967, three series of rechargeable single-cell silver-mercuric
oxide-zinc pacemakers have been implanted in dogs with complete heart
block. The five nonhermetic units in series 1 failed after less than or
equal to 18 months, primarily due to prototype cell deficiencies, although
one cell functioned for eight years. The six units in series II contained
improved cells, but failed due to gradual transepoxy fluid absorption after
less than or equal to 31 months. All rechargeable cells were salvaged and
dried, and, seven years after their manufacture, they continue to power
pacing circuits. Series III now totals 20 doubly hermetically sealed units,
tested for up to three years (total more than 300 months or 26 years), with
no pacemaker failures. Accelerated tests indicate a minimum life of more
than 50 years. A clinical trial is in progress.