A Method of Supporting Venous Tubing in the Neck to Prevent Occlusion of Peritoneovenous Shunts
William F. Sindelar, MD;
Steven T. Brower, MD
Arch Surg. 1984;119(3):349-350.
Abstract
Mechanical occlusion of the cervical venous tubing segment of a peritoneojugular shunt may occur because of kinking or angulation. This may be associated with partial or complete shunt failure. In four patients with inadequate palliation of ascites after shunt placement, restricted flow in the cervical tubing was diagnosed as the cause. When a stainless steel helical support was passed over the end of the tubing, angulation was eliminated and extended shunt patency was restored in all four patients. The helical support is recommended for use in patients who have or are at risk for the development of occlusion in the cervical venous tubing portion of the peritoneovenous shunt.
(Arch Surg 1984;119:349-350)
Author Affiliations
From the Surgery Branch, Clinical Oncology Program, Division of Cancer Treatment, National Cancer Institute, National Institutes of Health, Bethesda, Md.
Footnotes
Accepted for publication July 20, 1983.
Reprint requests to Surgery Branch, National Cancer Institute, National Institutes of Health, Bethesda, MD 20205 (Dr Sindelar).