Tactile Imaging of Breast Masses
First Clinical Report
Parris S. Wellman, PhD;
Edward P. Dalton, MD;
David Krag, MD;
Kenneth A. Kern, MD;
Robert D. Howe, PhD
Arch Surg. 2001;136:204-208.
Hypothesis Tactile imaging can accurately document the palpable extent of breast masses.
Design Prospective nonrandomized interventional trial, comparing mass size estimates from preoperative physical examination, ultrasound, and tactile imaging with postoperative measurements of the resected masses.
Setting A community ambulatory surgical center and a university hospital tertiary care center.
Patients Twenty-three women undergoing surgical excision of breast masses. All subjects had a single, palpable, dominant mass, 0.5 to 3 cm in diameter.
Intervention Prior to surgery, the size of each mass was estimated from tactile imaging using an array of pressure sensors that is stroked over the mass. Size was also estimated by ultrasound and physical examination. Immediately following resection of the mass, it was bisected, and the palpable extent was measured with a caliper.
Main Outcome Measure Maximum mass diameter estimates from ultrasound, physical examination, and tactile imaging, compared with the resected measurement.
Results Tactile imaging estimates were repeatable (7.5% mean SD for multiple estimates of the same mass) and show good agreement with the resected measurements. Mean absolute error was 13%, and linear regression with zero intercept had a slope of 0.94, r2 = 0.51. Physical examination and ultrasound estimates had respective mean absolute errors of 46% and 34%, regression slopes of 1.27 and 0.89, and r2 = 0.28 and 0.37.
Conclusions Tactile imaging can provide accurate and reproducible estimates of the size of breast masses. This capability can enhance cancer surveillance for patients with benign masses (eg, due to scarring or fibrocystic changes) because previous work suggests that reliable detection of a difference in mass size by physical examination requires a 40% change in diameter. In contrast, this study suggests tactile imaging requires only a 15% change (95% confidence interval).
From the Division of Engineering and Applied Sciences, Harvard University, Cambridge, Mass (Drs Wellman and Howe); New Hampshire Breast Center, Bedford (Dr Dalton); the Department of Surgery, University of Vermont, Burlington (Dr Krag); and the Department of Surgery, Hartford Hospital, Hartford, Conn (Dr Kern). A portion of salary support for this work was provided by the Whitaker Foundation, Rosslyn, Va (Drs Wellman and Howe), Drs Dalton, Krag, Kern, and Howe have served as consultants to Assurance Medical Inc.
CiteULike Connotea Del.icio.us Digg Reddit Technorati Twitter
What's this?
RELATED ARTICLE
This Month in Archives of Surgery
Arch Surg. 2001;136(2):143.
FULL TEXT
THIS ARTICLE HAS BEEN CITED BY OTHER ARTICLES
Differentiating Benign from Malignant Solid Breast Masses with US Strain Imaging
Burnside et al.
Radiology 2007;245:401-410.
ABSTRACT
| FULL TEXT
AAPM/RSNA Physics Tutorial for Residents: Topics in US: Beyond the Basics: Elasticity Imaging with US
Hall
RadioGraphics 2003;23:1657-1671.
ABSTRACT
| FULL TEXT
Concerns About Tactile Imaging and the Detection of Breast Masses
Goodson III
Arch Surg 2001;136:1327-1327.
FULL TEXT
|